Pre- and post-cruise calibration of optical instruments is highly recommended to help establish the measurements uncertainty. For example, some optical instruments – in particular, the ac-9, ac-S and backscattering and transmissometers with 660nm red LEDs – are known to drift significantly throughout a single cruise. It is therefore recommended, if high quality DIW is available